Technical data
— 14 SPECTROLAB — 30.07.2015
3 Technical data
Description Value
Instrument height 1405 mm; 55.3"
Instrument width 770 mm; 30.3"
Instrument depth 1900 mm; 74.8"
Instrument weight approx. 500 kg; ~ 1102 lbs
Power supply 198 - 253 VAC, 50/60 Hz
Power during measurement 1.0 kVA
Power during stand-by 0.5 kVA
Fuse 16 A (slow-blow)
Operating range (permissible
ambient temperature)
15°C – 30°C; 59°F – 86°F
Recommended temperature range
for reaching the specified
performance
18 – 30°C; 64 – 86°F
Operation Unless specified differently in the order, 2,500 m
max. above mean sea level.
Air humidity range 20 – 80%, not condensing
Pollution index 1
Sound pressure level 55 dB(A)
Stability (according to DIN EN
61010-1)
compliant if change of position <10°
Argon inlet pressure on the
instrument
5.0 bar
Argon quality (min.) 4.8 (99.998% Ar) mit O
2
<3ppm, N
2
<10ppm,
H
2
O<5ppm, kV<1ppm
When analysing for nitrogen/oxygen:
Argon > 99.999% with O
2
<0.5ppm, N
2
<0.5ppm;
H
2
O<3ppm, KW<1ppm
Argon flow rates Spark stand flushing
Stand-by (low flow)
5 L/h (LACM12)
Constant flow
30 L/h
Analytic flow
230 L/h (LAVM12)
140 L/h (LACM12)
Storage conditions Ambient temperature
-15 – 45°C; 5 – 113°F