6.4.2 Proceed with the measurement
To measure the probe, proceed as follows:
1. Start the required measuring program using function key F10.
2. Put the sample in position.
The sample must fully cover the opening in the spark stand plate.
The sample clamp must have contact with the sample (the safety circuit
is closed).
3. Use "Start" or the respective function key to start the measurement.
The measurement is fully automated.
The measurement is finished automatically.
The measurement results are displayed on the screen.
4. Perform further measurements if required.
Rotate the sample slightly for each further measurement. The respective
burn spots must not overlap.
5. Use "Finish" or press the respective function key to end the series of
measurement.
The measurement is complete.