Maintenance
— 34 SPECTROLAB — 30.07.2015
4. Turn the union nut (2) to remove
the exhaust hose from the spark
stand. Detach the hose from the
filter bottle.
5. Loosen the cable clamp on the
filter bottle.
6. Use compressed air or a similar
medium and carefully blow out
the exhaust hose.
7. Clean the connector (1) on the
spark stand.
8. Connect the exhaust hose to the
spark stand and filter unit.
Fig. 13
Cleaning of the exhaust hose is now complete.