Set up
HemoPoint
®
H2 User’s Guide
11
Place the photometer on a level counter adjacent to a power
socket. Make sure there is enough room for the cuvette holder to
be freely accessible.
Illustration 2: Setting up the photometer.
3.3 Switching the photometer on and off
3.3.1 Using for the first time
Allow the photometer to reach room temperature!
Changing from a cold to a warm environment (i.e. after storage or
transport), condensation can form both on the inside and the
outside of the photometer. Wait at least 1 hour, before you
connect the photometer to a power supply.
WARNING!