Troubleshooting
HemoPoint
®
H2 User’s Guide
52
7. Troubleshooting
Before you call Stanbio’s technical service department or send
the instrument in for repair, please try to define or solve the
problem with the help of this section.
Care! Danger of fatal electric shock!
It is essential to follow the instructions in Section 1.3 “Safety
notes”.
7.1 Problem solving
No display, no response to entries
Battery is empty Power the instrument with the power adaptor;
allow the battery to recharge completely.
Battery is defective Instrument can only be operated when
powered with the power adaptor.
Software does not
respond
Perform a reset of the instrument (see Section
7.2 “Resetting of the photometer”).
Battery problems
Error message:
"Defective battery"
A battery problem was found.
Following this message the battery is switched
off. The instrument can only be operated when
powered with the power adaptor.
->Contact Stanbio Technical Support.
No battery symbol
displayed
The battery has been switched off by the
instrument. The instrument can only be
powered with the power adaptor.
->Contact Stanbio Technical Support.
The battery operation
time is too short. The
instrument switches off
without previous
warning.
- Always leave the instrument connected to the
power adaptor until the charging process has
been completed.
- The battery has deteriorated and should be
replaced.
->Contact Stanbio Technical Support.