EasyManua.ls Logo

Stanbio HemoPoint-H2 - Setting Time Format; Setting the Power Frequency; Menu Options; Hematocrit Mode

Stanbio HemoPoint-H2
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Further functions
HemoPoint
®
H2 User’s Guide
40
5.3.7.3 Setting time format
Under “Menu”, sub-menu “Settings”, sub-menu “Time Format”
the output format of the time can be set. This affects the output
of data records in the memory (data storage) and the printout of
the results. The following settings are supported:
12 hours (default)
24 hours
5.3.7.4 Setting the power frequency
Under “Menu”, sub-menu “Settings”, sub-menu “Frequency” the
frequency of the electrical current can be set. In the United
States 60 Hz is the norm. The following settings are supported:
50 Hz
60 Hz (default)
The function of the individual keys can be found in Section 5.3.4
“Setting the units”.
5.3.8 Menu options
To keep the main menu clear, all functions that are used less
frequently have been placed under the menu item “Options”.
5.3.8.1 Hematocrit mode
Display of an estimated hematocrit value can be activated in
“Menu”, sub-menu “Options”, sub-menu “Hematocrit”. The
hematocrit value will be calculated using the formula Hct=F x
Hgb [g/dL], with F= 2.94. A true hematocrit test is not determined
with this system.
The use of this formula is allowed only within the normal
hemoglobin range, means from 12.0 g/dL (7.44 mmol/L) – 18.0
g/dL (11.16 mmol/L). If the Hgb result is outside this range then
the estimated hematocrit result will not be calculated and “N/A”
will appear.
WARNING!

Table of Contents