Maintenance
HemoPoint
®
H2 User’s Guide
47
6. Maintenance
6.1 Cleaning and disinfection of the instrument
6.1.1 Housing and touchscreen
It is essential to follow the instructions in Section 1.3 “Safety
notes”. Disconnect the power adaptor from the electrical
connection before proceeding.
Cleaning the housing and touchscreen is best accomplished with
a lint-free cloth, lightly dampened with clean water. For more
stubborn soiling, a mild soap solution may be used. For
disinfection, standard solutions can be used for surface
disinfection provided they do not
contain alcohol or other solvent.
The housing is equipped with a high-quality lacquered surface.
Please do not use any harsh cleaning agent or solvent otherwise
the surface can be damaged and the photometer will look
unsightly.
Before cleaning the touchscreen, open the cuvette holder.
6.1.2 Cuvette holder
The cuvette holder can be removed from the instrument for
cleaning. Proceed as follows:
Disconnect the power adaptor from the electrical
connection before proceeding.
Open the cuvette holder until you feel a resistance and
the holder will not extend further.
Press down the silver pin on the left-hand side (bottom)
of the cuvette holder with a ball-point pen and draw the
cuvette holder forward at the same time.
WARNING!
WARNING!
TIP