Further functions
HemoPoint
®
H2 User’s Guide
36
Model number of the photometer
- Version -
Software Version
Electronic hardware
Mechanical hardware
- Adaptation -
Correlation function to conform the output of results with different
devices and/or methods.
y = m * x + n
m … correlation factor
n … offset
if m = 1 and n = 0 (factory defaults) the function
is not active
The correlation can only be enabled by authorized personnel on
special request. Contact Stanbio Laboratory’s Technical Service
Department for additional information.
- SPN -
The current S
ervice Process Number, needed by Servicing
5.3.2 Setting the date
The function of setting the current date can be found under
“Menu”, sub-menu “Date”. Setting the date is important for the
correct relationships of the results from the memory (data
storage) or for printing out the results.
The format to enter the date depends on the date format setting
(see Section 5.3.7.2 “Setting date format”).
The arrow buttons increase or decrease
the numerical value in each case.
ESC
Press this button to reject the entry and to
leave the adjustment.