3. Dock the basket into the tank.
4. Put a solution of 5% non-foaming Sodium Hypochlorite disinfectant into a shallow
heat-resistant container (this prevents it from being drained at the start of the
cycle).
Disinfection solution Volume for Unity and Unity 20 tank
5% of non-foaming
Sodium Hypochlorite
5% is equivalent to
1750 ml (59.2fl.oz)
5. Put the container of disinfectant solution into the basket.
6. Close the lid and pull the lid handle towards you to engage the lid latches.
7. On the display, select the
Decontamination
wash program. Alternatively a custom
decontamination program can be defined using the supervisor menu if required.
8. Start the wash cycle.
72 MAINTENANCE
InnoWave Unity & Unity 20 Operator Manual