7-23
764333-675
7. Install new UV lamp by sliding into the UV chamber
slowly and rotating clockwise into the locked
position, as shown in Figure 7-27.
8. Install UV chamber lamp socket.
Press socket firmly and evenly onto the lamp pins
Align banana jack on lamp socket with slot on
the sleeve bolt
9. Lock the lamp socket using wire form. Pinch to lock
wire under locking tabs as shown in Figure 7-28.
10. Install shroud cover onto the standoffs on the
shroud plate using the pan head screws as shown
in Figure 7-29. After installation of a new lamp,
reset the elapse-time meter on the lamp power
supply. See Figure 7-29 and S
ECTION 7.27, RESET
UV LAMP ELAPSE-TIME METER.
Figure 7-27. Install New UV Lamp
Figure 7-28. Lock Lamp Socket with Wire Form
Figure 7-29. Install Shroud Cover