Live Document, Historical Table
Live
Document
Page
Number
Figure Number/ Name Manual Development Historical Record
(Front
Cover)
Front Cover
P: Stamp "01/27/11 LIVE DOCUMENT" C: Document Manufacturing
changes/manual errors and clarifications A: Stamped to show Live Document Date
4-3
Table 4-1 Preventive
Maintenance Schedule
P: Frequency of 3.1, Replacement of Check Valves CK1, CK2, CK3 and CK11 is
incorrect.
C: Frequency changed.
A: Step 3.1, Check Valves CK1, CK2, CK3 and CK11 need to be replaced 2x/year.