Operating the Device
28
EN
5.13 Launching/Configuring the User Menu
Device parameters are changed and displayed with the user menu.
Hold down the menu key (MENU) and turn on the device using the ON/OFF key.
Keep the (MENU) key depressed until the LED test is finished. The nominal flow
display will depict a P after the LED test and emit an acoustic signal. The device
has now entered the first menu level of the menu mode.
The first menu level offers a choice between User menu and Service menu. The
nominal pressure display depicts P for User menu or L for Service menu
(accessible only to service technicians).
Switch from one to the other with the / keys.
Confirm with the (MENU) key.
Select the menu level or the value to be adjusted using the / keys.
The values to be modified are displayed in the nominal flow display.
•Adjusted values are saved with the (MENU) key (use like ENTER key).
Exit the menu or menu level without saving with the Start/Stop key
functioning as the ESC key.
To reenter OP mode, turn the device off and on again.
The following user menu levels and settings are available:
Parameter Description Factory
Settings
P User menu (accessible to the user)
L Service menu (accessible only to service personnel)
P01 Volume level of pump unit alarm.
Two possible settings (1=muted, 2=loud).
2
P02
Setting, if balancing unit connected or not.
0 = not connected
1 = connected
first saved, then turn on again.
After saving the next menu points are not longer accessible.
1
P03
Setting for the deficit threshold - acoustic warning levels in the balancing unit
Setting range 0-2000 ml.
When pressing the key briefly -> in increments of 10
When pressing the key longer -> in increments of 100
Additional depiction in the deficit display
750 ml
P04 Volume of the acoustic warning in the balancing unit
Two possible settings (1=muted, 2=loud).
Additional depiction in the deficit display.
2
P05 Check of the mass gauge of the balancing unit
Display in windows for deficit (0) – and nominal flow (0.00).
Nominal
Pressure
80 mmHg
Nominal
Flow
150 ml/min