Error and Warning Messages
44
EN
13 Error and Warning Messages
Signals Error and Cause Solution / Device Response
Acoustic alarm, 3x short Pressure chamber of tube set not
properly positioned.
Check position of pressure chamber.
Listen for audible click indicating
correct fit of tube set pressure
chamber.
Acoustic alarm for duration of
exceeding pressure parameter
Actual pressure display flashes
Nominal pressure has been
exceeded->Actual pressure has
exceeded permissible tolerance value
of 10 mm Hg for longer than 5 seconds.
The electronics of the pump unit
automatically reduces the pressure
by reversing the pump motor. The
duration of the reverse movement of
the roller wheel is limited to 9
seconds. Electronic control system
automatically turns off the pump
motor. Pressure must be externally
reduced. Once pressure value falls
bellow the nominal pressure, pump
motor turns on again.
Actual pressure display flashes
No acoustic alarm
Overpressure alarm 150 mm Hg,
Device maximal pressure 150 mm Hg
has been reached. Actual pressure
equals or exceeds 150 mm Hg.
•None
Acoustic alarm, 1x long A safety threshold message is
displayed starting at 100 mm Hg.
Release the key for one second to
set values between 100 to 150 mm
Hg. Use the same key subsequently
to set the pressure up to 150 mm Hg.
Acoustic alarm for duration of
exceeding pressure parameter
Actual pressure display flashes
Icon for overpressure is lit.
Overpressure alarm 200 mm Hg
Actual pressure equaled or exceeded
200 mm Hg for longer than 5 seconds.
Electronic control system of pump
unit automatically turns off the
pump motor and turns it back on
once value falls below 200 mm Hg.
Acoustic alarm, 3x short
Deficit value flashes
The deficit threshold set by the
operator has been reached.
•None
Acoustic alarm, 3x short
Deficit value flashes
Alarm indicates value is 100 ml
above the deficit threshold;
additional increase of the deficit
after exceeding the deficit threshold
every 100 ml.
•None
Acoustic alarm, 3x short
Deficit value flashes
Deficit rate icon is lit
Acoustic alarm 5x short, rapid
Deficit rate alarm
The fluid deficit is equal or greater
than 300 ml/min.
•None
Acoustic alarm 5x short, rapid
ID flashes in the actual pressure
display for 5 seconds
Instrument detection has failed.
Pressure values during the instrument
detection are not plausible.
Automatic stop and switching back
of the device to the irrigation
pressure display.
To pump fluid, press the Start/Stop
key again or repeat the instrument
detection sequence as described in
chapter 5.10 Instrument Detection,
page 25.
Acoustic alarm 2x long
SCALE flashes in the actual pressure
display for 5 seconds
Scale overloaded, the load capacity
value of 42 kg was exceeded for 30 s.
Remove load from weighing unit
The overload counter in the device is
set.