Error and Warning Messages
45
EN
E01
Service icon is lit
Continuous acoustic alarm
E01 flashes in the actual pressure
display.
Pressure measuring failure
/malfunction.
Pressure sensor 1 and/or 2 defective.
Replace the device when error occurs
permanently.
E02
Service icon is lit
Continuous acoustic alarm
E02 flashes in the actual pressure
display
Failure of the weighing unit / display
unit Data lines or weighing unit
defective.
Replace device.
Pump without balancing still to be
used for diagnosis.
E50
Service icon is lit
Acoustic alarm, 1x
E50 flashes in the actual pressure
display
Communication error
Malfunctioning communication
between modules.
Replace device.
E51
Service icon is lit
Acoustic alarm, 2x
E51 flashes in the actual pressure
display
Communication error
Electronics error
Replace device.
E59
Acoustic alarm, 10x
E59 flashes in the actual pressure
display
Safe status, only M key is active
Self-test calibration error
Initial calibration not executed or
interrupted.
•New calibration required.
No display in the display unit. Scale connection not activated in the
user menu.
Activate the scale connection in the
user menu item P02.