12 Technical Data
Power connection: 100-240 V~
Frequency: 50/60 Hz
Mains fuse: 2x T3, 15 A, UL-recognized
Max. power consumption: 82 VA
Max. current consumption: 100 V: 880 mA
240 V: 340 mA
Protection class: I
Protection level: BF
Moisture protection: IP41 (Pump unit)
Moisture protection: IP21 (Weight unit)
Classification: IIb
Grounding: Potential equalization plug
The device complies with the
following standards:
IEC 60601-1:1998 +A1:1991+A2:1995
IEC 60601-1:2005+Corr2006+Corr2007+Int1:2008+Int2:2009
EMC: EN 60601-1-2 / IEC 60601-1-2
Operating conditions: 10-40 °C / 50-104 °F
30 to 75% rel. humidity
Storage and transportation
conditions:
-40 to +70 °C /-40 to +158 °F
10 to 90% rel. humidity
Dimensions: Width x height x depth
710 x 1400 x 810 [mm]
28 x 55 x 32 [inch]
Mass: 26 kg /10.36 lb
Max. weight: 70 kg /10.36 lb
Load capacity: 0-40 kg
Max. load of weighing system: 50 kg
Maximum output pressure: 200 mm Hg
Adjustable pressure range: 15 to 100 and 100 to 150 mm Hg
Accuracy of pressure measurement: ± 1,5 measure value
Max. flow: 500 ml/min
Accuracy of flow measurement: +10%/-3% measure value
±3% measure value
Accuracy of deficit display: ±10ml, ±3% of capacity, at 21 °C/ 69.8 °F
Connections/ports: Data inputs/outputs