40 6082-209-001 REV A www.stryker.com
Return To Table of Contents
Preventative Maintenance
Operation Schedule Procedure
Cleaning and Disinfecting Each Use. See page 38
Inspection
For 1-25 calls per month, inspect cot
every 6 months
For 26-200 calls per month, inspect
cot every 3 months
For 200+ calls per month, inspect cot
monthly.
See below for checklist
Note: Keep up to date maintenance records using the Maintenance Record form on page 43.
CHECKLIST
_____ All fasteners secure (reference all assembly drawings)
_____ All welds intact, not cracked or broken
_____ No bent or broken tubing or sheet metal
_____ No debris in wheels
_____ All wheels secure, rolling and swiveling properly
_____ Optional wheel lock holds wheel securely when on and clears wheel when off
_____ Siderails move an latch properly
_____ Backrest operating properly
_____ Optional accessories intact and operating properly
_____ Height positioning latch functioning properly
_____ Cot secure in each height position
_____ Undercarriage folds properly
_____ Breakaway head section operating properly
_____ Safety bar operating properly
_____ Foot rest operating properly
_____ No rips or cracks in mattress cover
_____ Body restraints intact and working properly
_____ Lubricate base tubes (optional)
Serial Number:
Completed by: _______________________________________ Date: _________________

Table of Contents

Other manuals for Stryker MX-PRO R3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker MX-PRO R3 and is the answer not in the manual?

Stryker MX-PRO R3 Specifications

General IconGeneral
BrandStryker
ModelMX-PRO R3
CategoryMedical Equipment
LanguageEnglish

Related product manuals