EasyManua.ls Logo

Stryker MX-PRO R3 - Preventative Maintenance

Stryker MX-PRO R3
125 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40 6082-209-001 REV A www.stryker.com
Return To Table of Contents
Preventative Maintenance
Operation Schedule Procedure
Cleaning and Disinfecting Each Use. See page 38
Inspection
For 1-25 calls per month, inspect cot
every 6 months
For 26-200 calls per month, inspect
cot every 3 months
For 200+ calls per month, inspect cot
monthly.
See below for checklist
Note: Keep up to date maintenance records using the Maintenance Record form on page 43.
CHECKLIST
_____ All fasteners secure (reference all assembly drawings)
_____ All welds intact, not cracked or broken
_____ No bent or broken tubing or sheet metal
_____ No debris in wheels
_____ All wheels secure, rolling and swiveling properly
_____ Optional wheel lock holds wheel securely when on and clears wheel when off
_____ Siderails move an latch properly
_____ Backrest operating properly
_____ Optional accessories intact and operating properly
_____ Height positioning latch functioning properly
_____ Cot secure in each height position
_____ Undercarriage folds properly
_____ Breakaway head section operating properly
_____ Safety bar operating properly
_____ Foot rest operating properly
_____ No rips or cracks in mattress cover
_____ Body restraints intact and working properly
_____ Lubricate base tubes (optional)
Serial Number:
Completed by: _______________________________________ Date: _________________