www.stryker.com 6082-209-001 REV A 41
Return To Table of Contents
Preventative Maintenance
BASE LUBRICATION
Note: The MX−PRO
®
has been designed to operate without the need for lubrication. Tri−Flow™ with Teflon
®
lubricant
may be used to reduce the force required to raise the undercarriage and to minimize wear. Do not use silicone,
WD−40™, or lithium grease. They could harm the moving parts of the cot.
1. Place the cot in the highest position.
2. Turn the cot upside down with the base legs facing up.
3. Using the plastic applicator tube, spray Tri−Flow™ with
Teflon
®
lubricant (Stryker part number 6082−19912) in
the gap between the upper (item A) and lower lift tubes
(item B) on both foot end legs.
4. Apply Tri−Flow™ to right and left height adjustment
Height Adjustment Racks racks providing an even
application.
5. Apply Tri−Flow™ to the right and left slide plates
providing an even application.
6. Let the cot sit for 5 minutes.
7. Turn the cot back over. Raise and lower the cot a
few times to work the lubricant throughout the applied
areas.
8. Wipe any excess lubricant from the base tubes. Slide
Plates
B
A
FOOT END
APPLY LUBRICATION
Figure 37 - Base Lubrication Locations
Height Adjustment Racks
Slide Plates

Table of Contents

Other manuals for Stryker MX-PRO R3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker MX-PRO R3 and is the answer not in the manual?

Stryker MX-PRO R3 Specifications

General IconGeneral
BrandStryker
ModelMX-PRO R3
CategoryMedical Equipment
LanguageEnglish

Related product manuals