Troubleshooting
PROBLEM CAUSE ACTION
Docking Operation
The rover will not dock or an error occurs during
docking.
The docker is not receiving facility water. Ensure water inlet hose is connected correctly. Ensure the
facility water supply valve is open. Dock the rover.
Communication between the rover and docker is
interrupted.
Pull the rover away from the docker. Remove and apply power
to the docker. Wait 60 seconds. Dock the rover.
Docker does not dispense detergent during the
cleaning cycle.
Bottle of detergent is empty. Replace the bottle of the detergent. See the label instructions
on the Stryker-approved detergent.
The end of the detergent inlet tube is not immersed
in the detergent.
Push the end of the detergent inlet tube to the bottom of the
detergent bottle.
The detergent inlet tube is not connected securely
to the detergent inlet port of the docker.
Connect the detergent inlet tube to the detergent inlet port of
the docker securely.
An error occurs during docking and the rover
cannot be removed from docker.
The fluid connectors prevent the removal of the
rover from the docker.
Remove and apply power to the docker. Wait 10 seconds. Pull
the rover away from the docker.

Other manuals for Stryker Neptune 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Neptune 2 and is the answer not in the manual?

Stryker Neptune 2 Specifications

General IconGeneral
BrandStryker
ModelNeptune 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals