4 www.stryker.com
EN 0702-002-700 Rev-J
For Use With
The Stryker Rover (rover) is for use with the Neptune 2 Waste Management
System, specifically the Neptune 2 Docking Station (docker).
Indications For Use
The Neptune Waste Management System is intended to be used in the
operating room, pathology, surgical centers, and doctor’s offices to collect
and dispose of surgical fluid waste as well as collect smoke generated from
electrocautery or laser devices.
Contraindications
None known.
Description
The rover is a mobile unit used to suction and collect fluid waste and small
debris from a surgical site in an operating room. Mobility allows the rover to be
relocated to a waste disposal area where the rover’s collection canisters may
be emptied.
Fluid waste is removed from the surgical site through suction tubes connected
to inlet ports of manifold(s) installed in the rover. The fluid waste and small
debris are collected in the canister(s) of the rover. The two canister design
allows separate canister suction level measurement and volume measurement
capability. However, both canisters use a single vacuum pump. A suction level
decrease in either canister, due to an open port, may cause the suction level
in the other canister to decrease. After collection, the rover is relocated to a
docking station located in a waste disposal area. The rover interfaces with the
docking station to empty the rover’s canister(s) of fluid waste. After disposal,
the rover canister(s) are rinsed with clean water and Stryker-approved
detergent, if desired, to facilitate cleaning.
The rover also has a height-adjustable, powered IV pole and a smoke
evacuator.
Important Information
The words WARNING, CAUTION and NOTE have special meaning and
should be reviewed.
WARNING:
Disregarding WARNING information may compromise the
safety of the patient and/or healthcare staff and may result
in injury.
CAUTION:
Disregarding CAUTION information may compromise
product reliability and may result in damage.
NOTE:
NOTE information supplements and/or clarifies procedural
information.
A triangle with an exclamation point alerts the healthcare
professional to read and understand the accompanying
instructions, especially the operating, maintenance, and
safety information.
*If you need more information or a complete list of accessory information, contact your
Stryker sales representative or call Stryker Neptune customer service at 1-800-550-7836.
Outside the US, contact your nearest Stryker subsidiary.
Accessories*
WARNING: Use only Stryker-approved components and
detergent, unless otherwise specified. Using other components
may result in increased electromagnetic emissions or decreased
electromagnetic immunity of the system. DO NOT modify any
component or accessory.
Description REF
Neptune Disposable Manifolds .............................................0702-02X-XXX Series
Neptune Fluid Suction HEPA Filter .....................................................0702-034-000
Neptune Smoke Evacuator ULPA Filter ..............................................0702-040-000
Smoke Tubing, 3/8 inch x 10 feet .......................................................0702-045-023
Smoke Tubing, 7/8 inch x 10 feet .......................................................0702-045-025
PenAdapt
TM
........................................................................................... 0702-045-027
PenAdapt
is the trademark of Buffalo Filter.
Compliance Statements
NOTE: The stated Federal Communications Commission (FCC) and Industry
Canada (IC) information applies only to Rover units equipped with an enabled
Stryker RF Module REF 0702-014-028. Contact Stryker regarding the RF
Module feature and its availability.
Federal Communications Commission (FCC)
FCC ID: Q9R-0702014028
This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions: (1) This device may not cause harmful interference,
and (2) this device must accept any interference received, including
interference that may cause undesired operation.
NOTE: FCC regulations provide that changes or modifications not expressly
approved by Stryker Instruments could void your authority to operate this
equipment.
Industry Canada (IC)
IC: 4919A-014028
The term "IC:" before the radio certification number only signifies that Industry
Canada technical specifications were met.
This equipment complies with FCC and IC radiation exposure limits set forth for
an uncontrolled environment. This equipment is in direct contact with the body
of the user under normal operating conditions. This transmitter must not be co-
located or operating in conjunction with any other antenna or transmitter.