CODE MESSAGE ACTION
0.X SYSTEMERROR CALL SERVICE
1. X MEMORYERROR CALL SERVICE
2.2 SYSTEMERROR CALL SERVICE
3.0 DOCKER COMM
ERROR
REMOVEROVER-CYCLEDOCKER
POWER - WAIT 60 SECONDS - DOCK
ROVER AGAIN
3.1 COUPLING ERROR REMOVEROVER-CYCLEDOCKER
POWER - WAIT 60 SECONDS - DOCK
ROVER AGAIN
3.2 COUPLING ERROR CALL SERVICE
3.3 OFFLOAD ERROR REMOVE MANIFOLDS FROM CANISTERS -
DOCK ROVER AGAIN
3.4 OFFLOAD ERROR CALL SERVICE
3.5 PREFILL ERROR CHECKFACILITYWATERSUPPLY-DOCK
ROVER AGAIN
3.6 PREFILL ERROR CHECKFACILITYWATERSUPPLY-DOCK
ROVER AGAIN
3.7 COUPLING ERROR CALL SERVICE
3.9 COUPLING ERROR CALL SERVICE
3.12 DOCK INCOMPLETE DOCK ROVER AGAIN
3.13 DOCKERNOTREADY REMOVEROVER-CYCLEDOCKER
POWER - WAIT 60 SECONDS - DOCK
ROVER AGAIN
4.0 LARGE CAN FULL SMALL CANISTER VOLUME WILL NOT FIT
IN LARGE - DOCK ROVER WHEN ABLE
4.1 VACUUM ERROR - SM CALL SERVICE
4.2 EMPTYTANKERROR CALL SERVICE
4.3 EMPTYTANKERROR DOCK ROVER TO REMOVE DEBRIS FROM
SMALL CANISTER
4.4 PREFILL ERROR PREFILLTANKISEMPTY-DOCKROVER
4.5 PREFILL ERROR PREFILLTANKISEMPTY-DOCKROVER
5.0 FILTER EXPIRED REPLACE FLUID SUCTION HEPA FILTER
5.1 FILTER EXPIRED REPLACE SMOKE FILTER
6.0 EMPTYTANKERROR CALL SERVICE
6.1 EMPTYTANKERROR CALL SERVICE
7.0 LEVEL SENSOR
ERROR
CALL SERVICE
7.1 LEVEL SENSOR
ERROR
CALL SERVICE
7.2 LEVEL SENSOR
ERROR
CALL SERVICE
7.7 PREFILL ERROR CHECKFACILITYWATERSUPPLY-DOCK
ROVER AGAIN
7.8 PREFILL ERROR CHECKFACILITYWATERSUPPLY-DOCK
ROVER AGAIN
CODE MESSAGE ACTION
8.X LEVEL SENSOR ERROR CALL SERVICE
9.0 IV POLE ERROR CALL SERVICE
9.1 IV POLE ERROR REDUCE POLE LOAD WEIGHT
9.2 IV POLE ERROR WAIT10MIN-RETRYPOLE
10.0 SMOKE EVAC ERROR CALL SERVICE
10.1 SMOKE EVAC ERROR CALL SERVICE
10.2 NO SMOKE FILTER SMOKE FILTER NOT DETECTED -
REPLACE SMOKE FILTER
11.X VOLDISPLAYERROR CALL SERVICE
12.0 LEVEL SENSOR ERROR CALL SERVICE
12.1 LEVEL SENSOR ERROR CALL SERVICE
12.2 LEVEL SENSOR ERROR CALL SERVICE
12.4 TEMP SENSOR ERROR CALL SERVICE
12.5 TEMP SENSOR ERROR CALL SERVICE
12.6 LEVEL SENSOR ERROR CALL SERVICE
12.7 LEVEL SENSOR ERROR CALL SERVICE
13.0 VACUUM ERROR - LG CALL SERVICE
13.1 VACUUM ERROR - LG CALL SERVICE
13.2 VACUUM ERROR - LG CALL SERVICE
13.3 VACUUM ERROR - LG CALL SERVICE
13.4 VACUUM ERROR - LG CALL SERVICE
13.5 VACUUM ERROR - LG CALL SERVICE
13.6 VACUUM ERROR - LG CANISTER IS PRESSURIZED
EXTERNALLY-TURNONVACUUM
SYSTEM
13.7 VACUUM ERROR - LG CALL SERVICE
13.8 VACUUM ERROR - LG CALL SERVICE
13.10 VACUUM ERROR - LG CALL SERVICE
13.11 VACUUM ERROR - LG CALL SERVICE
14.X VACUUM ERROR - LG CALL SERVICE
15.1 VACUUM ERROR - SM CALL SERVICE
15.2 VACUUM ERROR - SM CALL SERVICE
15.3 VACUUM ERROR - SM CALL SERVICE
15.4 VACUUM ERROR - SM CALL SERVICE
15.5 VACUUM ERROR - SM CALL SERVICE
15.6 VACUUM ERROR - SM CANISTER IS PRESSURIZED
EXTERNALLY-TURNONVACUUM
SYSTEM
15.7 VACUUM ERROR - SM CALL SERVICE
15.8 VACUUM ERROR - SM CALL SERVICE
15.10 VACUUM ERROR - SM CALL SERVICE
15.11 VACUUM ERROR - SM CALL SERVICE
16.X VACUUM ERROR - SM CALL SERVICE