16 www.stryker.com
EN 0702-002-700 Rev-J
Specifications
1
Model:
Neptune 2 Rover-ULTRA
REF:
0702-001-000 0702-002-000
Electrical Power Requirements:
120 V
~
, 60 Hz, 12.0 A, single phase
20 V
,
3 A during docking procedure; rover
receives power from docking station
230 V
~
, 50 Hz, 9.0 A, single phase
20 V
,
3 A during docking procedure; rover
receives power from docking station
Size:
19inch[48.3cm]width
23inch[58.4cm]depth
102inch[259cm]height(IVpoleup)
70inch[177.8cm]height(IVpoledown)
Vacuum Range: 50to530mm-Hg[2to21in-Hg,7to71kPa](measuredwithallportsclosed)
Vacuum Measurement:
+ 5% of full scale (+ 1.05 in-Hg)
Weight:
300lbs.[136kg]-collectionempty
353lbs.[160kg]-collectionfull
Volume:
24-liter capacity (combination of 4 and 20 liter canisters)
Volume Measurement:
2
4-liter canister - + 50 mL
20-liter canister - + 150 mL
Inclined Plane of Operation:
+ 2.5 degrees
Equipment Type:
Class I, Type CF Applied Part
Enclosure Protection:
IPX0 Ordinary Equipment
Infrared Communication Windows:
Protective Earth Ground:
Approval:
CSA International
CAN/CSA-C22.2 No. 601.1 M90: 2003
BS EN 60601-1: 1997
UL 60601-1: 2003
AS/NZS 4268: 2008
N 17693
Environmental Conditions: Operation Storage and Transportation
(before initial use)
Storage and Transportation
(after initial use)
Temperature Limitation:
Humidity Limitation:
Atmospheric Pressure Limitation:
1
Specifications are approximate and may vary from unit to unit or by power supply fluctuations.
2
Volume measurements specified do not account for fluid evaporation or an inclined plane of operation that exceeds the specified range.