EasyManua.ls Logo

Swann AMI INSPECTOR pH - Connecting Sample Inlet and Outlet

Swann AMI INSPECTOR pH
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AMI INSPECTOR pH
Product Description
A-96.250.761 / 161017 11
2.3. Technical Data
Power Supply Battery
Use original, supplied power adapter only.
Voltage:
Power consumption:
Charging Time:
Battery type:
85265 VAC, 50/60 Hz
max. 20 VA
~ 6h
Li-Ion
During charging protect from heat impact and keep splash-proof
(not IP66).
Measuring
Range
pH:
Resolution:
1 to 12 pH
0.01 pH
Temperature
Input
for Pt1000 type sensor
Measuring range:
Resolution:
- 30 to +130 °C
0.1 °C
Sample
conditions
Flow Rate:
Temperature:
Inlet pressure:
5–10 l/h
up to 50 °C
0.22 bar
Temperature
monitoring
Alarm if the transmitter temperature is higher than + 65 °C or lower
than 0 °C.