12 A-96.250.761 / 161017
AMI INSPECTOR pH
Installation
3. Installation
3.1. Installation Checklist
Check
Instrument’s specification must conform to your AC power
ratings. See External power adapter, p. 20.
Check if the battery is fully charged.
Installation
Connect sample inlet and outlet to the flow cell.
pH electrode
Install the sensor (see Install Swansensor pH SI, p. 14).
Connect sensor cables.
Store the protective caps for later use.
Power-up
Turn on the sample flow and wait until the flow cell is
completely filled.
Switch on power.
Instrument
set-up
Adjust sample flow.
Program all parameters for sensor.
Program all parameters for instrument operation (limits,
alarms).
Run-in period
Let the instrument run continuously for 1 h.