EasyManua.ls Logo

Swann AMI INSPECTOR pH - Changing Parameters and Values

Swann AMI INSPECTOR pH
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
28 A-96.250.761 / 161017
AMI INSPECTOR pH
Operation
5.4. Changing Parameters and values
Changing
parameters
The following example shows how to change the logger interval:
Changing
values
1 Select the parameter you want to
change.
2 Press <Enter>
3 Press < > or< > to high-
light the required parameter.
4 Press <Enter> to confirm the se-
lection or <Exit> to keep the previ-
ous parameter).
The selected parameter is
highlighted (but not saved yet).
5 Press <Exit>.
Yes is highlighted.
6 Press <Enter> to save the new pa-
rameter.
The system reboots, the new
parameter is set.
5.1.2
Sensors
Sensor type FOME
Temperature NT5K
Standards
Disinf. Free chlorine
4.4.1
Logger
Log interval 30 min
Clear logger no
4.1.3
Logger
Clear logger no
Log interval 30min
1 Hour
Interval.
5 min
30 min
10 min
4.1.3
Logger
Log interval 10 min
Clear logger no
4.1.3
Logger
Log interval event driven
Clear logger no
No
Save ?
Yes
1 Select the value you want to
change.
2 Press <Enter>.
3 Set required value with < >
or< >.
4 Press <Enter> to confirm the new
value.
5 Press <Exit>.
Yes is highlighted.
6 Press <Enter> to save the new val-
ue.
5.3.1.1.1
Alarm High 12.00 pH
Alarm
Alarm Low -3.00 pH
Hysteresis 0.10 pH
Delay 5 Sec
5.3.1.1.1
Alarm
Alarm Low -3.00 pH
Hysteresis 0.10 pH
Delay 5 Sec
Alarm High 9.00 pH

Related product manuals