Maintenance
A-96.250.761 / 161017 39
6.6. Replacing Fuses
WARNING
External Voltage.
External supplied devices connected to relay 1 or 2 or to the
alarm relay can cause electrical shocks.
Make sure that the devices connected to the following con-
tacts are disconnected from the power before resuming in-
stallation.
relay 1
relay 2
alarm relay
Find and repair the cause for the short circuit before replacing the
fuse. Use tweezers or needle-nosed pliers to remove the defective
fuse.
Use original fuses provided by SWAN only.
A 1.25 AF/250V Instrument power supply