30 A-96.250.761 / 161017
AMI INSPECTOR pH
Maintenance
Clean
pH Electrode
1 If necessary wipe the electrode shaft and the green tip cautious-
ly with a soft, clean, and damp paper tissue.
2 Remove grease with a tissue moistened with alcohol.
3 Slightly open the sensor cap of reference electrode and let flow
out about 5 ml of electrolyte.
4 Rinse the electrode tip thoroughly with clean water.
5 Tighten the sensor cap hand tight.
6 Install the electrode into the flow cell again.
7 Let the electrode run-in for 1 h before the first calibration.
A
B
Connector
Electrode
C
D
Union screw
Flow cell block
B
A
C
D
A
B
Sensor cap tightened
Sensor cap slightly opened
AB