36 A-96.250.761 / 161017
AMI INSPECTOR pH
Maintenance
1 Stop sample flow to the monitor AMI pH-Redox by closing the
appropriate valve, e.g. back pressure regulator, sample prepa-
ration or flow regulating valve at flow cell.
2 Connect sample line of the Monitor AMI pH-Redox [A] with the
sample inlet of the reference instrument AMI INSPECTOR pH.
Use the supplied tube, made of FEP.
3 Connect sample outlet of the reference instrument AMI IN-
SPECTOR pH to the sample outlet funnel of the monitor.
4 Switch on AMI INSPECTOR pH. Open the flow regulating valve
and regulate the sample flow.
A
B
C
D
Monitor AMI pH-Redox
KCl reservoir
pH electrode
Temperature sensor
E
F
G
Flow cell block
Sample inlet
Sample outlet
A
B
C
D
E
F
G