AMI INSPECTOR pH
Error List
A-96.250.761 / 161017 43
Error Description Corrective action
E001 Alarm high
check process
check programmed value, see
5.3.1.1.1, S. 59
E002 Alarm low
check process
check programmed value, see
5.3.1.1.26, S. 59
E007 Sample Temp. high
check process
check programmed value, see
5.3.1.3.1, S. 60
E008 Sample Temp. low
check process
check programmed value, see
5.3.1.3.26, S. 60
E009 Sample Flow high
check sample flow
check programmed value, see
5.3.1.2.2, S. 60
E010 Sample Flow low
establish sample flow
clean instrument
check programmed value, see
5.3.1.2.36, S. 60
E011 Temp. shorted
Check wiring of temperature sensor
Check temperature sensor
E012 Temp. disconnected
Check wiring of temperature sensor
Check temperature sensor
E013 Case Temp. high
check case/environment temperature
check programmed value, see 5.3.1.4,
S. 60
E014 Case Temp. low
check case/environment temperature
check programmed value, see 5.3.1.5,
S. 60
E017 Control Timeout
check control device or programming in
Installation, Relay contact, Relay 1/2
see 5.3.2/3, S. 61
E024 Input active
See If Fault Yes is programmed in
Menu see 5.3.4, S. 65