AMI INSPECTOR pH
Error List
A-96.250.761 / 161017 43
Error Description Corrective action
E001 Alarm high
– check process
– check programmed value, see
5.3.1.1.1, S. 59
E002 Alarm low
– check process
– check programmed value, see
5.3.1.1.26, S. 59
E007 Sample Temp. high
– check process
– check programmed value, see
5.3.1.3.1, S. 60
E008 Sample Temp. low
– check process
– check programmed value, see
5.3.1.3.26, S. 60
E009 Sample Flow high
– check sample flow
– check programmed value, see
5.3.1.2.2, S. 60
E010 Sample Flow low
– establish sample flow
– clean instrument
– check programmed value, see
5.3.1.2.36, S. 60
E011 Temp. shorted
– Check wiring of temperature sensor
– Check temperature sensor
E012 Temp. disconnected
– Check wiring of temperature sensor
– Check temperature sensor
E013 Case Temp. high
– check case/environment temperature
– check programmed value, see 5.3.1.4,
S. 60
E014 Case Temp. low
– check case/environment temperature
– check programmed value, see 5.3.1.5,
S. 60
E017 Control Timeout
– check control device or programming in
Installation, Relay contact, Relay 1/2
see 5.3.2/3, S. 61
E024 Input active
– See If Fault Yes is programmed in
Menu see 5.3.4, S. 65