60 A-96.250.531 / 300617
AMI Oxytrace QED
Error List
Error Description Corrective action
E001 Oxygen Alarm high
– check process
– check programmed value,
see 5.3.1.1.1, S. 79
E002 Oxygen Alarm low
– check process
– check programmed value,
see 5.3.1.1.25, S. 79
E003 Saturation Alarm high
– check process
– check programmed value,
see 5.3.1.4, S. 80
E004 Saturation Alarm low
– check process
– check programmed value,
see 5.3.1.4, S. 80
E007 Sample Temp. high
– check process
– check programmed value,
see 5.3.1.3.1, S. 80
E008 Sample Temp. low
– check process
– check programmed value,
see 5.3.1.3.25, S. 80
E009 Sample Flow high
– check sample flow
– check programmed value,
see 5.3.1.2.2, S. 80
E010 Sample Flow low
– establish sample flow
– clean instrument
– check programmed value,
see 5.3.1.2.35, S. 80
E011 Temp. shorted
– Check wiring of sensor
– Check sensor
E012 Temp. disconnected
– Check wiring of sensor
– Check sensor