AMI Oxytrace QED
Error List
A-96.250.531 / 300617 61
E013 Case Temp. high
– check case/environment temperature
– check programmed value,
see 5.3.1.5.1, S. 80
E014 Case Temp. low
– check case/environment temperature
– check programmed value,
see 5.3.1.5.2, S. 80
E017 Control Timeout
– check control device or programming in
Installation, Relay contact, Relay 1/2
see 5.3.2 and 5.3.3, S. 81
E018 Faraday Efficiency
– clean Faraday electrode, see
Maintenance of the Faraday electrode,
S. 45
– perform an air calibration, see
Calibration, S. 47
– perform maintenance of the oxygen
sensor, see Maintenance of the Oxygen
Sensor, S. 42
E019 Quality Assurance
– Perform QA Procedure using a
reference instrument, e.g. AMI Inspector
E024 Input active
– See If Fault Yes is programmed in Menu
see 5.3.4, S. 85
E026 IC LM75
– call service
E028 Signal output open
– check wiring on signal outputs 1 and 2
E030 EEProm Frontend
– call service
E031 Calibration Recout
– call service