Swann AMI Powercon - Error Codes and Corrective Actions

Swann AMI Powercon
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AMI Powercon
Error List
A-96.250.341 / 061015 63
Error Description Corrective action
E001 Cond. Alarm high
check process
check programmed value, see 5.3.1.1,
p. 79
E002 Cond. Alarm low
check process
check programmed value, see 5.3.1.1,
p. 79
E007 Sample Temp. high
check process
check programmed value, see 5.3.1.3,
p. 80
E008 Sample Temp. low
check process
check programmed value, see 5.3.1.3,
p. 80
E009 Sample Flow high
check sample inlet pressure
check programmed value, see
5.3.1.2.2, p. 80
E010 Sample Flow low
check sample inlet pressure
Check flow regulating valve
check programmed value, see
5.3.1.2.35, p. 80
E011 Temp. shorted
Check wiring of temperature sensor
Check temperature sensor
E012 Temp. disconnected
Check wiring of temperature sensor
Check temperature sensor
E013 Case Temp. high
check case/environment temperature
check programmed value, see 5.3.1.4,
p. 80
E014 Case Temp. low
check case/environment temperature
check programmed value, see 5.3.1.5,
p. 80
E017 Control time-out
Check control device or programming
in Installation, Relay contact, Relay 1/2
5.3.2/3, p. 81
E018 Quality Assurance
Perform QA Procedure using reference
instrument, e.g. AMI Inspector