AMI Powercon
Error List
A-96.250.341 / 061015 63
Error Description Corrective action
E001 Cond. Alarm high
– check process
– check programmed value, see 5.3.1.1,
p. 79
E002 Cond. Alarm low
– check process
– check programmed value, see 5.3.1.1,
p. 79
E007 Sample Temp. high
– check process
– check programmed value, see 5.3.1.3,
p. 80
E008 Sample Temp. low
– check process
– check programmed value, see 5.3.1.3,
p. 80
E009 Sample Flow high
– check sample inlet pressure
– check programmed value, see
5.3.1.2.2, p. 80
E010 Sample Flow low
– check sample inlet pressure
– Check flow regulating valve
– check programmed value, see
5.3.1.2.35, p. 80
E011 Temp. shorted
– Check wiring of temperature sensor
– Check temperature sensor
E012 Temp. disconnected
– Check wiring of temperature sensor
– Check temperature sensor
E013 Case Temp. high
– check case/environment temperature
– check programmed value, see 5.3.1.4,
p. 80
E014 Case Temp. low
– check case/environment temperature
– check programmed value, see 5.3.1.5,
p. 80
E017 Control time-out
– Check control device or programming
in Installation, Relay contact, Relay 1/2
5.3.2/3, p. 81
E018 Quality Assurance
– Perform QA Procedure using reference
instrument, e.g. AMI Inspector