6 Measurement with the Autoloading Station
CyFlow
Robby 6 Autoloading Station | Operating Manual | September 2020 | Revision 002
15
using the Sample Port (manual mode) before starting Sample Analysis with delivery from
the Autoloader tray (Automate mode). The other measure modes deliver samples from
the Autoloader tray of the Robby 6 Autoloading Station and define different ranges of
sample volume.
Figure 8: Changeable parameters of work step
Within this window it is necessary to place the “Device” selection to “AL 1” → the different
conditions of the measure modes are only then visible and speed and/or analysis volume
settings are changeable.
The analysis volume and the analysis speed for the respective volume range can be
selected in the field “Changeable parameters” (see below).
NOTICE
The sample uptake will be higher than the analysis volume.
Sample uptake
(fixed)
Analysis volume
(User selectable)
For 96-well-plate
50 µl
10 40 µl
120 µl
40 100 µl
200 µl
100 180 µl
For tubes
400 µl + X
X = 10 2000 µl
Table 2: Volume ranges of the Robby 6 Autoloading Station