7 Cleaning Procedure
CyFlow
Robby 6 Autoloading Station | Operating Manual | September 2020 | Revision 002
21
7.3 Sheath Fluid and Waste bottle
CAUTION
Direct contact with biohazardous material
Possibly biohazardous material like samples and Waste bottle contents
may cause severe illness.
Strict guidelines, international, as well as national regulatory standards
such as PPE (personal protection equipment) must be met for all
users.
The Waste bottle content must be discarded accordingly to the relevant biohazard
regulations.
A regular thorough cleaning of the Sheath Fluid bottle and exchange of the Inline Filter
will keep the background in the measurements to a minimum level.
Sheath Fluid bottle can be cleaned with Sheath Fluid or Hypochlorite Solution 0.5%
(Order No. 04-4012)