7 Cleaning Procedure
20
7 Cleaning Procedure
7.1 Autoloader tray
The sample needle of the CyFlow
Robby 6 Autoloading Station will be cleaned
automatically between each sample and at the end of each measuring sequence.
7.2 Sample Port
The sample port of the CyFlow
Ploidy Analyser can be cleaned using the Cleaning
Solution (green solution, Order No 04-4009_R) followed by a Sample Tube with Sheath
Fluid. This procedure allows to reduce significantly the cross contamination and reduce
the background.
For this procedure, connect a Sample Tube with 1.5 ml Cleaning Solution and clean the
instrument by clicking on the “CLEAN” symbol .
The instrument will perform one cleaning cycle. Repeat the procedure with a Sample
Tube filled with 1.5 ml Sheath Fluid.