Direct contact with biohazardous material
Possibly biohazardous material like samples and Waste bottle contents
may cause severe illness.
• Strict guidelines, international, as well as national regulatory standards
such as PPE (personal protection equipment) must be met for all
users.
Clean the housing of the CyFlow
™
Robby 6 Autoloading Station and its components on a
regular basis carefully with soft cloth. Water must not enter the CyFlow
™
Robby 6
Autoloading Station or peripheral devices or come into contact with electric connections
and switches. For cleaning the screen, always use special screen cleaner and soft cloth.
Do not use any organic solvents, nitro thinner, benzol, alcohol, highly concentrated
bleach etc.!
Regularly empty the Waste bottle and clean with warm detergent solution and a brush.
Clean Sheath Fluid reservoir with distilled water and a clean brush and flush with clean
distilled water several times.
Remember: a clean Sheath Fluid reservoir is critical for proper operation.
If the CyFlow
™
Robby 6 Autoloading Station will not be used for longer periods, clean flow
system by using distilled water. Put a sample tube half-way filled with distilled water on
the sample port. Clean waste and Sheath Fluid reservoir, wipe top dry.
8.2 Preventive maintenance and service
Installation, Service and Preventive Maintenance are all part of the authorised service
personnel responsibilities, please contact your local Sysmex representative.
If you have any questions about the content of this operating manual or the use of the
device, contact your local Sysmex representative.
8.3 Transport and Storage
For the transport of the system to a different location it will be necessary to disconnect all
external data and supply connections. The system should be carried in upright position.
During transport or storage please take care, the system will be stored under the following
conditions: