Document Ref: SUKBMS-24-
601
Version: 2.0
Page | 12
Date: 19/07/2021
Default smear settings
Level
HCT %
Sample
vol
(pulse)
Angle
(degrees)
Speed
(mm/sec)
Standby
time
(sec)
Smear
start
pos.
(pulse)
Lower
Limit
Upper
limit
1
-
20
13
27.0
150
2.0
100
2
20
25
13
27.0
140
2.0
100
3
25
30
13
27.0
130
2.0
100
4
30
35
13
27.0
120
2.0
100
5
35
40
13
27.0
110
2.0
100
6
40
45
13
27.0
100
2.0
100
7
45
50
13
27.0
90
2.0
100
8
50
-
13
27.0
80
2.0
100
Once smear is complete slide is moved back by the slide glass shift unit (D) to the smear drier unit
(4) where they are dried. From here they are moved to the slide print position (5) where the printer
(F) prints sample information on the frosted end of the slide. From the slide print position (5) the
slide is moved through a seconded drier to ensure both smear and print are dried to the slide
storage position (7). A single slide cassette moves in from the cassette supply table to the single
cassette rotator (9) where it is rotated through 90º to allow the prepared smear to be moved into it
at the slide storage position (7). The single cassette contain the prepared smear is then rotated
back through 90º to be transported to the smear table (10).