TA I
NSTRUMENTS
TGA 2950 1–13
Specifications
tt
tt
t
CAUTION:
Table 1.4
Sampling System
Sample pans
Types Platinum, Alumina
(Al
2
0
3
), Aluminum
Volume capacity Platinum: 50 µL,
100 µL
Alumina: 100 µL,
250 µL, 500 µL
Aluminum 100 µL
Weighing capacity
1
1.0 gm
Balance measurement
2
Resolution 0.1 µg
Accuracy
< + 0.1%
Ranges 100 mg range:
0.1 µg
_
100 mg
1000 mg range:
1 µg
_
1000 mg
1
The total mechanical
capacity of the balance is
5 gm. In order to avoid
damaging the balance
assembly, never allow the
total weight of the sample,
tare weight, hang-down
wires, and pans to exceed
5 gm.
2
The TGA balance mechanism is sensitive
to changes in the surrounding room tempera-
ture. For optimum accuracy, you must
regulate the ambient temperature.
(table continued)