Appendix B
B–62 TA I
NSTRUMENTS
TGA 2950
Varying
Sensitivity Setting
In Figure B.14 we have overlaid the 1
°
C/minute
conventional constant heating rate decomposi-
tion of the mixture (curve a) with a dynamic rate
Hi-Res
TM
scan at four different sensitivity
settings (curves b through e). All of the Hi-
Res
TM
scans were run at 20
°
C/minute with a
resolution setting of 5.0. Note that increasing
sensitivity setting increases the sharpness of
each transition, but does not substantially
change the transition temperature.
Figure B.14
(a) 1
°
C/min Normal Ramp; (b) Sensitivity setting 1; (c) Sensitivity
setting 2; (d) Sensitivity setting 4; (e) Sensitivity setting 8.