TA I
NSTRUMENTS
TGA 2950
Installing the 2950
2–22
Installing the
Hang-Down Wires
During installation, take care not to bend the
hang-down wires or damage the hang-down loops.
1. Turn on the TGA instrument.
2. Press the FURNACE key to lower the
furnace.
3. Locate the sample hang-down wire in your
TGA Accessory Kit.
4. Hold the wire in your hand so that the
doubly bent top hook is pointing to the left
and the bottom hook is pointing to the right.
5. Carefully insert the bottom of the hangdown
wire into the top of the furnace far enough
so that you can insert the top of the wire into
the thermocouple tube without bending the
wire (Figure 2.14).
Figure 2.14
Installing the Sample
Hang-Down Wire
t
CAUTION: