Introducing the TGA 2950
TA I
NSTRUMENTS
TGA 29501–6
NOTE:
A pound sign (#) after the weight signal indi-
cates that the balance reading has not yet
stabilized. When the weight stabilizes, the
pound sign will disappear.
2950 Keypad
The instrument keypad (see Figure 1.2) contains
the keys found in Table 1.1 and the HEATER
and POWER switches
Experiment information and instrument constants
are entered from the controller keyboard, not the
instrument keypad.
Table 1.1
TGA 2950 Keypad
Function Keys
Key/Function Explanation
SCROLL Scrolls the realtime
signals shown on the
bottom line of the display.
For more details on the
experiment, refer to status
and signal displays on the
controller.
TARE Zeros the displayed
weight of an empty
sample pan: automatically
loads the pan from the
sample platform, raises
the furnace to protect the
pan from air currents,
weighs the pan, stores the
weight as an offset, and
then unloads the pan.
(table continued)