Appendix C
TA I
NSTRUMENTS
TGA 2950C–10
Calibrating
the Auto TGA
When using the Auto TGA accessory, the need
for calibration remains the sameweight
calibration and temperature calibration are
recommended at least once a month or when
replacing the thermocouple; and the platform
adjustment procedure needs to be performed if
the sample hang-down wire fails to pick up a
sample pan. See the online help and documenta-
tion for further information.