TGA 2950 25
this position. While your assistant holds the
unit, remove the two nuts and washers from
the other side.
6. Slide the unit completely onto the lab bench.
Have your assistant hold one side up while
you unscrew and remove the black rubber
shipping feet from the bottom. Then rotate
the unit and remove the shipping feet from
the other side in the same manner.
7. Have your assistant lift the entire unit while
you slide the plywood board out from under
the unit.
8. Have your assistant lift one side of the unit
while you use a wrench to install two
mounting feet on the other side (see Figure
2.3). Rotate the unit and install the two
remaining mounting feet in the same man-
ner.
9. Remove the furnace clamp before turning on
the power to the unit.
Figure 2.3
Installing the
Mounting Feet
Unpacking/Repacking the 2950