TGA 2950 2–11
Figure 2.5
Heat Exchanger
Bottle
2. Add TA Instruments TGA Conditioner (PN
952377.901) into the water reservoir bottle.
Refer to the instructions on the bottle for the
amount of conditioner to add to the reser-
voir. Then fill the bottle to the inner rim
(see Figure 2.5) with distilled water.
After the system has been started, recheck the
level of water in the reservoir bottle and refill to
the inner rim if necessary.
Do not put any liquid other than distilled
water in the heat exchanger reservoir.
3. Replace and tighten the water reservoir cap.
Installing the Instrument
Fill to bottle’s inner rim.