Model C606Important: To the Operator
120515
Faults Occurring While in AUTO Mode
(L/R) HPR>41F (5C) AFTER 4 HR - The mix
temperature in the left or right hopper was above
41°F (5°C) more than four hours.
(L/R) BRL>41F (5C) AFTER 4 HR - The mix
temperature in the left or right barrel was above
41°F (5°C) more than four hours.
(L/R) HPR>45F (7C) AFTER 1 HR - The mix
temperature in the left or right hopper was above
45°F (7°C) more than one hour.
(L/R) BRL>45F (7C) AFTER 1 HR - The mix
temperature in the left or right barrel was above
45°F (7°C) more than one hour.
(L/R) HPR>41F (5C) AFTER PF - The mix
temperature in the left or right hopper was above
41°F (5°C) more than four hours following a power
failure.
(L/R) BRL>41F (5C) AFTER PF - The mix
temperature in the left or right barrel was above
41°F (5°C) more than four hours following a power
failure.
(L/R) HPR>59F (15C) - The mix temperature in the
left or right hopper exceeded 59°F (15°C).
(L/R) BRL>59F (15C) - The mix temperature in the
left or right barrel exceeded 59°F (15°C).
The FAULT HISTORY screens will display up to 100
faults that have occurred. The most recent fault is
displayed on screen 1. The date, time, and fault
description is displayed on each screen.
(See Figure 75.)
FAULT HISTORY 1
05/15/12 08:00
FAULT DESCRIPTION
>Exit
Figure 75
Advance to the next most recent occurring fault by
selecting the AUTO symbol
. Scroll the screens in
the opposite direction by selecting the OPTIONAL
FLAVOR symbol
.ExittheFAULTHISTORY
screen and return to the menu by selecting the
CALIBRATION symbol
.
Fault Descriptions
(L/R) Comp On Too Long - The left or right main
compressor has run for more than 11 consecutive
minutes without dispensing product.
(L/R) Product Door Off - The left or right freezer
door is not completely installed or the safety
interlock circuit has opened.
(L/R) Hopper Therm Bad - The left or right hopper
thermistor probe is SHORTED or OPEN.
(L/R) Hopper Over Temp - The left or right hopper
thermistor probe is reading over 200F(93C).
(L/R) Barrel Over Temp - The left or right barrel
thermistor probe is reading over 200F(93C).
(L/R) Beater Overload - The left or right reset
mechanism has tripped.
(L/R) HPCO Compressor - The left or right high
pressure switch contacts have opened.
(L/R) Glycol Therm Bad - The left or right glycol
thermistor probe is reading over 200F(93C).
The HEAT CYCLE SUMMARY screen displays the
hours since the last heat cycle, the hours since the
product temperature was above 150°F (65.6°C), and
the number of heat cycles completed since the last
brush clean date.
HEAT CYCLE SUMMARY
HRS SINCE HC 0
HRS SINCE 150 0
HC SINCE BC 0
Figure 76
The HEAT CYCLE DATA screen contains a record
of up to 366 heat treatment cycles. The most recent
heat cycle data will be shown first. The Standard
records have each heat cycle recorded in three
screens. Select the AUTO symbol
to move the
arrow to Standard records“ and select the
CALIBRATION symbol
. (See Figure 77.)
HEAT TREAT CYCLE
> Standard records
Details
Exit
Figure 77