ATP Operator Instructions
|Selecting an Installation Site
13
Chapter 3 | SELECTING AN INSTALLATION SITE
Overview
This unit is a precision instrument. To operate this unit properly and to obtain maximum performance,
install it in a location that meets the following conditions. The unit stores a certain amount of medications
in the tablet canisters while dispensing the required amount. Temperature and humidity conditions must
be closely monitored since medications are temporarily stored.
Install the unit indoors.
Select a location not exposed to direct sunlight and that is free from dust.
Install it on a sturdy, level floor.
Select a location with low humidity. But, avoid a location that is too dry.
Do not install it near a wash basin or kitchen sink.
Install it in a location without stoves, gas ranges and other heat-generating equipment.
Do not install it near equipment that generates strong magnetic fields.
Select a location away from TV or radio sets. Otherwise, the unit may distort the picture and sound of these
appliances.
Leave a space of at least 10 cm from the rear side of the unit to ensure that the internal fan is not blocked.
Ambient temperature:15 to 28
Ambient humidity: 20.to 70 %R.H.
At an altitude up to 6,562 feet (2000 m)
Using a mains supply voltage that does not exceed 115VAC ± 10%, 60Hz. <6 amperes, with transient over
voltages II in accordance with IEC-60664-1
Consistent with Pollution Degree 2 in accordance with IEC-60664-1
Use a grounded AC wall outlet to prevent electric shock. Have authorized personnel install a grounded AC
wall outlet in a location that is not provided with one.