Page #170 background imageLoading...
Page #170 background image
Appendix
C
Table C-1
Confidence Test Numbers
and
Affected
Functions
Test
Number Description
01
Controller Timing Signal Missing or Wrong Period
02 Momentary Switch Stuck
03
Readout Interface or Memory Failure
04
Calibration Data Parity or Checksum Error
05
Main Board Failure Detected by Auto Level Trigger on LINE
Source
06
Memory-Battery Voltage Too High or Too Low
GP
11
GPIB Interface Failure Detected
OM
76
Digital Multimeter Failure Detected
CT
81
- 87 Counter-Timer-Trigger Failure Detected
C-2
2465A/2455A/2445A Operators

Table of Contents