Gain setting, full scale, ATI channel DPO77002SX
DPO75902SX, DPO75002SX
100 mV 1.19 mV
200 mV 1.76 mV
250 mV 2.10 mV
300 mV 2.49 mV
Input resistance
≤1.2 V
FS
settings 50 Ω ±3% at 18 to 28 ºC (64 to 82 ºF)
50 Ω ±4% over 5 to 45 ºC (45 to 113 ºF)
>1.2 V
FS
settings 50 Ω ±4.4% over 5 to 45 ºC (45 to 113 ºF)
ATI channel 50 Ω ±3% from 18 °C to 28 °C
50 Ω ±4% from 5 °C to 45 °C
Horizontal and acquisition system
Long term sample rate accuracy ± 0.1 ppm initial accuracy. Aging < 0.8 ppm first year, <0.3 ppm/year after first year. Applies only
when using the internal reference.
Vertical system analog channels (cont.) Specifications
DPO70000SX Series User 191