EasyManua.ls Logo

Tektronix PS 5010

Tektronix PS 5010
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration-PS
501
0
Internal Adjustment Procedure
The
DVM
must be connect-
ed directly across the
0.1
Q
resistor sense leads. The
sense leads break out the
side of the resistor body.
3391
-20
PS
5010
LOGIC
Fig.
5-7.
Test setup for
Max
I
Adj., step
2.
DVM
3.
Adjust
+
and
-
Da
0
Adj
a. Turn power module off.
u
b.
Move the J1320 jumper to the calibration position. Its
location is shown in Fig. 6-1.
NOTE
J1320
remains
in
the calibration
positon
for the re-
maining adjustments.
Connect the dvrn negative lead to TP1330, the Ref
Com test point.
Connect the dvrn positive lead to TP1431, the V
+
Dac
test point.
Turn the power module on
Program the positive and negative supplies for 0 V
output.
g.
Make certain the OUTPUT front panel button is
illuminated.
h. CHECK-that the dvrn reads from -0.20
mV to
+0.20 mV.
i.
ADJUST-R1335,
+
Da 0 Adj, for a dvrn reading of
-
0.20 mV to +0.20 mV.
j.
Move the dvrn positive lead to TP 1430, the V-Dac test
point.
k.
CHECK-for a dvrn reading from -0.20 mV to
+
0.20 mV.
I.
ADJUST-R1334, -Da 0 Adj, for a dvrn reading from
-0.20 mV to +0.20 mV.
m. Remove these connections for the next step.
Adjust
+O
Adj
Connect the dvrn
+
lead to the floating supply
POSITIVE connector and the dvrn -lead to the front
panel common (ground) connector.
Make certain the positive supply is programmed for 0 V
output, and the OUTPUT is ON.
CHECK-for a dvrn reading from -1.00 mV to
+
1 .OO mV.
ADJUST-R1345,
+O
Adj, for a dvrn reading of
-
1 .OO mV to
+
1
.OO
mV.
Remove the connection to the front panel POSITIVE
terminal for the next step.
Adjust
-0
Adj
Connect the dvrn positive lead to the PS 5010
NEGATIVE terminal and the dvrn negative lead to the
front
ane el
common terminal.

Table of Contents

Related product manuals