Page #63 background image
DETAILED
COMMAND
LIST
I
DT
(device trigger)
Type:
Setting or query
Setting syntax:
DT SET
DT OFF
Examples:
DT SET
DT OFF
Query syntax:
DT?
Query response:
DT SET;
DT
OFF;
Discussion:
This command causes the instrument to wait for the
<GET> (group execute trigger) interface message before
updating the hardware to new settings. The OFF argument
permits hardware updating without the <GET> message.
ERROR?
Type:
Query
Query syntax:
ERR:
Query ERROR? response:
ERR
103;
ERR
204;
Discussion:
This command returns an error code for the most recent
error reported via serial poll. When the error status reporting
is disabled
(RQS
OFF)
this command returns the highest
priority condition pending. The condition is cleared and not
reported again.
ERROR?