FSOUTPUT
Type:
Setting or query
Setting syntax:
FSOUT ON
FSOUT OFF
Examples:
FSOUT ON
FSOUTPUT OFF
Query syntax:
FSOUT?
Query response:
FSOUT ON;
FSOUT OFF;
Discussion:
This command connects or disconnects both floating
supplies to or from their respective output terminals. Some
protective components remain connected to the output ter-
minals. Refer to schematics
11
and 12 for components that
remain connected.
FSOUTPUT
Type:
Query
Query syntax:
ID?
.
Query response:
ID
TEKIPS501 01V79.1 ,FXX;
Discussion:
XX indicates the firmware version number.
ID?